Application feasibility of a near-infrared closed-eye pupil monitoring system in awake neurosurgical patients
Author:
Affiliation:

Fund Project:

  • 摘要
  • |
  • 图/表
  • |
  • 访问统计
  • |
  • 参考文献
  • |
  • 相似文献
  • |
  • 引证文献
  • |
  • 资源附件
    摘要:

    目的 评估基于近红外技术的闭眼瞳孔监测系统用于神经外科患者的安全性、可行性以及测量性能。 方法 招募神经外科清醒患者30例进行前瞻性验证研究,以睁眼状态下人工手动测量值作为临床对照标准,评价闭眼瞳孔监测系统的安全性和可行性,结合平均绝对误差、均方根误差和Bland-Altman一致性分析来评价测量准确性。 结果 安全性验证研究中,受试者在系统移除设备后10 min内(T1)及72 h内(T2)随访均未记录到设备相关不良事件和物理安全性事件。该系统在闭眼状态下的自动测量值与睁眼人工参考值相比,平均绝对误差为0.14 mm,均方根误差为0.16 mm。Bland-Altman一致性分析显示,平均偏倚为-0.10 mm,95%一致性界限为-0.36~0.15 mm。 结论 基于近红外技术的闭眼瞳孔监测系统用于神经外科清醒的住院患者具有较好的安全性、可行性及可靠的测量性能。

    Abstract:

    Objective To assess the application safety, feasibility, and measurement performance of a near-infrared system that monitors pupil size through closed eyelids in neurosurgical patients. Methods We prospectively enrolled 30 awake neurosurgical patients in this prospective validation study.Manual pupil measurements obtained with the eyes open were used as the clinical reference standard.We assessed the safety and feasibility of the closed-eye pupil monitoring system and evaluated its measurement accuracy using mean absolute error, root mean square error, and Bland-Altman analysis. Results In the safety validation study, no device-related adverse events or physical safety incidents were recorded at either follow-up assessment, conducted within 10 minutes of device removal (T1) and within 72 hours (T2).Compared with the manual reference measurements obtained with the eyes open, the automated closed-eye measurements had a mean absolute error of 0.14 mm and a root mean square error of 0.16 mm.Bland-Altman analysis showed a mean bias of -0.10 mm, with 95% limits of agreement ranging from -0.36 to 0.15 mm. Conclusion Application of the near-infrared closed-eye pupil monitoring system in awake neurosurgical inpatients is safe and feasible, and it exhibits reliable measurement performance.

    参考文献
    相似文献
    引证文献
引用本文

商芷颖,李景宇,梁双,张令玮,柏晓燕,徐博,陈璐.近红外闭眼瞳孔监测系统用于神经外科清醒患者的可行性研究[J].护理学杂志,2026,41(17):33-36+42

复制
文章指标
  • 点击次数:
  • 下载次数:
历史
  • 收稿日期:2026-04-19
  • 最后修改日期:2026-06-22
  • 录用日期:
  • 在线发布日期: 2026-10-09