Abstract:Objective To compare the clinical effectiveness of three-position dynamic ultrasound with single-position static ultrasound for accurate catheter tip localization during peripherally inserted central catheter (PICC) placement in neonates. Methods A prospective comparative study with a time-sequence design was conducted. A total of 128 neonates admitted to the neonatal intensive care unit who underwent PICC placement were consecutively enrolled and allocated into two groups according to the chronological period of catheter insertion, with 64 participants in each group. In the control group, bedside ultrasound with a single position (functional position) was used for static assessment and localization of the catheter tip during PICC placement. In the intervention group, three-position bedside ultrasound (adduction position, functional position, and abduction position) was used to dynamically assess the catheter tip, and the catheter tip position was adjusted until the optimal location was achieved. Results The incidence of primary catheter malposition in the intervention group (7.81%) was significantly lower than that in the control group (21.88%, P<0.05). In addition, the intervention group had significantly fewer radiographic examinations, and catheter repositioning procedures than the control group (all P<0.05). Conclusion Three-position dynamic ultrasound enabled dynamic and precise localization of the PICC catheter tip in neonates, effectively reducing catheter malposition and related complications. It also decreased radiation exposure, and improved the safety of PICC placement and nursing care in neonates.